Liquid Filled Drip Proof Thermometer M Series
Hyoda (Japan) Waterproof Bimetal Thermometer M series
Model: M3, M4, M6
Anritsu HD series are high accuracy handheld thermometers by Anritsu meter. The high accuracy is realized by adopted the thermocouple material from input plug to cold junction It has automatic standard voltage calibration system, hence achieved high accuracy, high response and high stability. This is the right tool if you are looking for precise temperature measurement!
|
HD series handheld thermometers allow easy temperature measurement. These general-purpose thermometers can be used for on-site temperature measurement and various other applications. |
|
Basic model
|
Standard model
|
High function model
|
LED Display model
|
Memory model
|
Basic model
|
Standard model
|
Model | HD-1100 | HD-1200 | HD-1201 | HD-1202 | HD-1300 | HD-1301 | HD-1302 | HD-1400 | HD-1500 | ||
Display | LCD with Backlight | LCD with Backlight | LCD with Backlight | LCD with Backlight | LCD with Backlight | LCD with Backlight | LCD with Backlight | LED | LCD with Backlight | ||
Waterproof specifications | IPX5 | IPX5 | - | - | - | - | - | IPX5 | - | ||
Input connector | ASP | ||||||||||
Input type | Thermocouple Input: Type E, K (Not switchable) | ||||||||||
Channel | 1ch | ||||||||||
Signal source resistance | MAX.500ohm | ||||||||||
Measurement Range | 1°C Resolution | E | -200~800°C | ||||||||
K | -200~1370°C | ||||||||||
0.1°C Resolution | E | -104.9~504.9°C(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.) | |||||||||
K | -104.9~504.9°C(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.) | ||||||||||
Measurement Accuracy | 1°C Resolution | 0°C~ | ±(0.1% of reading +1°C) | ||||||||
~0°C | ±(0.5% of reading +1°C) | ||||||||||
0.1°C Resolution | 0°C~ | ±(0.05% of reading + 0.2°C) | |||||||||
~0°C | ±0.5°C | ||||||||||
Reference junction compensation accuracy | ±0.2°C at 25°C±10°C | ||||||||||
Temperature Coefficient | ±0.01% of F/S per °C (F/S : full scale) | ||||||||||
Operation Environment | 0 to 40°C, 0 to 80%RH (Non condensing) 32 to 104°F, 0 to 80%RH (Non condensing) |
||||||||||
Storage Environment | -20 to 50°C, 0 to 85%RH (Non condensing) -4 to 122°F, 0 to 85%RH (Non condensing) |
||||||||||
Battery life | 300h | 300h | 150h | 150h | 300h | 150h | 150h | 150h | 150h | ||
Power supply(Dry battery) | 4 AA Alkali Batteries | ||||||||||
Power supply(AC power) | - | - |
AC-Adaptor (Option) |
- |
AC-Adaptor (Option) |
||||||
Sampling rate | approx.300ms |
approx. 300ms *1 |
|||||||||
Linearlizer | Digital Linearizer method (Compliant with IEC 60584-1 (2013)) | ||||||||||
Dimensions(mm) | approx. 76(w)×167(H)×36(D) [Protruded parts excluded] | ||||||||||
Weight | approx. 350g [Include Batteries] | ||||||||||
Supplied accessories | Manual,Test report,Soft case, Hand strap,4 AA Alkaline batteries | ||||||||||
Optional accessories | - | - | Analog output Cable | Alarm output Cable | - |
Communi cation Cable |
|||||
- | Analog output Cable |
Software (AMS-100) |
|||||||||
Analog Output | - |
10mV/°C (table1) |
- |
1mV/°C (table1) |
10mV/°C (table1) |
- | - | ||||
Alarm Output | - | - | - |
(table2) |
(table2) |
- | - | ||||
Memory function | - | - | - | - | - | - |
(table3) |
||||
Compatible standards | CE Marking, RoHS |
*1 Sampling period will be a user-selected time interval when time interval is specified at 10 seconds or more. |
Model |
HD-1150 HD-1650 |
HD-1250 HD-1750 |
HD-1251 | HD-1252 | HD-1350 | HD-1351 | HD-1352 | HD-1450 | HD-1550 | ||
Display | LCD with Backlight | LCD with Backlight | LCD with Backlight | LCD with Backlight | LCD with Backlight | LCD with Backlight | LCD with Backlight | LED | LCD with Backlight | ||
Waterproof specifications | IPX5 | IPX5 | - | - | - | - | - | IPX5 | - | ||
Input connector | ANP | ||||||||||
Input type | Thermocouple Input: Type E, K (Not switchable) | ||||||||||
Channel | 1ch | ||||||||||
Signal source resistance | MAX.500ohm | ||||||||||
Measurement Range | 1°C Resolution | E | -200~800°C | ||||||||
K | -200~1370°C | ||||||||||
0.1°C Resolution | E | -104.9~504.9°C(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.) | |||||||||
K | -104.9~504.9°C(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.) | ||||||||||
Measurement Accuracy | 1°C Resolution | 0°C~ | ±(0.1% of reading +1°C) | ||||||||
~0°C | ±(0.5% of reading +1°C) | ||||||||||
0.1°C Resolution | 0°C~ | ±(0.05% of reading + 0.2°C) | |||||||||
~0°C | ±0.5°C | ||||||||||
Reference junction compensation accuracy | ±0.2°C at 25°C±10°C | ||||||||||
Temperature Coefficient | ±0.01% of F/S per °C (F/S : full scale) | ||||||||||
Operation Environment | 0 to 40°C, 0 to 80%RH (Non condensing) 32 to 104°F, 0 to 80%RH (Non condensing) |
||||||||||
Storage Environment | -20 to 50°C, 0 to 85%RH (Non condensing) -4 to 122°F, 0 to 85%RH (Non condensing) |
||||||||||
Battery life | 300h | 300h | 150h | 150h | 300h | 150h | 150h | 150h | 150h | ||
Power supply(Dry battery) | 4 AA Alkali Batteries | ||||||||||
Power supply(AC power) | - | - |
AC-Adaptor (Option) |
- |
AC-Adaptor (Option) |
||||||
Sampling rate | approx.300ms |
approx. 300ms *1 |
|||||||||
Linearlizer | Digital Linearizer method (Compliant with IEC 60584-1 (2013)) | ||||||||||
Dimensions(mm) | approx. 76(w)×167(H)×36(D) [Protruded parts excluded] | ||||||||||
Weight | approx. 350g [Include Batteries] | ||||||||||
Supplied accessories | Manual,Test report,Soft case, Hand strap,4 AA Alkaline batteries | ||||||||||
Optional accessories | - | - | Analog output Cable | Alarm output Cable | - |
Communi cation Cable |
|||||
- | Analog output Cable |
Software (AMS-100) |
|||||||||
Analog Output | - | - |
1mV/°C (table1) |
10mV/°C (table1) |
- |
1mV/°C (table1) |
10mV/°C (table1) |
- | - | ||
Alarm Output | - | - | - | - |
(table2) |
(table2) |
(table2) |
- | - | ||
Memory function | - | - | - | - | - | - | - | - |
(table3) |
||
Compatible standards | CE Marking, RoHS |
*1 Sampling period will be a user-selected time interval when time interval is specified at 10 seconds or more |
HD-1201 HD-1251 |
HD-1301 HD-1351 |
HD-1202 HD-1252 |
HD-1302 HD-1352 |
|
Rate | 1mV/°C | 10mV/°C | ||
Range | All range at 1°C resolution | All range at 0.1°C resolution | ||
Resistance | 100ohm | |||
Accuracy | ± (0.15% of F/S + 1mV) at 25°C±5°C± (0.15% of F/S + 1mV) ±(0.01% of F/S) except 25°C±5°C |
Output Pattern | Upper Limit | Lower Limit |
More than upper limit | H(5V) | L(0V) |
Between upper and lower limit | L(0V) | (0V) |
Less than lower limit | L(0V) | H(5V) |
|
Time interval | second:1, 5, 10, 30 minutes: 1, 5, 10, 30, 60 and manual memory |
Memory capacity | 9999 data |
Available models:
|