The Vaisala HMT140 Wi-Fi Data Logger has been discontinued and is no longer recommended for new monitoring systems.
The HMT140 was a battery-powered Wi-Fi data logger for use with the Vaisala viewLinc Continuous Monitoring System. Depending on the model, it could measure relative humidity, temperature, RTD inputs, voltage, current or Boolean contacts.
This page remains available to help existing users identify installed HMT140 models, maintain operating systems and plan migration to current Vaisala monitoring hardware.
| Existing HMT140 Application | Recommended Direction |
|---|---|
| Temperature and relative humidity | RFL100 with AP10 |
| Refrigerator, freezer or cold-room temperature | RFL100 with a suitable temperature probe |
| CO₂, temperature and humidity | RFL100 with compatible probes and AP10 |
| Voltage, current or Boolean input | Application review required; a VDL200 or another viewLinc-compatible interface may be more suitable |
The RFL100 uses Vaisala VaiNet rather than the HMT140’s direct Wi-Fi connection. Each installation therefore requires an AP10 Access Point and a check of wireless coverage, probe compatibility and viewLinc version.
| Legacy Model | Measurement Channels |
|---|---|
| HMT141 | 1 temperature and 1 relative-humidity channel using an HMP110 probe |
| HMT143 | 2 RTD temperature channels |
| HMT144 | 2 voltage inputs, 0–5 V DC |
| HMT145 | 2 voltage inputs, 0–10 V DC |
| HMT146 | 2 current inputs, 0–20 mA |
| HMT147 | 2 Boolean or door-contact inputs |
| HMT148 | 1 RTD temperature input and 1 Boolean input |
| HMT14D | 1 RTD temperature input and 1 current input |
| HMT14E | 1 RTD temperature input and 1 voltage input, 0–5 V DC |
| HMT14F | 1 RTD temperature input and 1 voltage input, 0–10 V DC |
| Measurement | Legacy Specification |
|---|---|
| Relative humidity | 0–100% RH using the HMP110 probe |
| HMP110 temperature | −40 to +80 °C |
| RTD temperature input | −196 to +90 °C |
| Current input | 0–22 mA measurement range |
| Voltage input | 0–5 V or 0–10 V, depending on model |
| Contact input | Open/closed dry contact |
| Wireless network | IEEE 802.11b/g Wi-Fi |
| Wi-Fi security | WEP 128-bit, WPA or WPA2 Personal |
| Memory | 3,060 samples |
| Sample interval | User selectable from 2 to 60 minutes |
| Display | Optional 128 × 64 monochrome display without backlight |
| Housing protection | IP65 |
| Operating temperature without display | −40 to +60 °C |
| Operating temperature with display | −20 to +60 °C |
| Weight | 300 g with batteries, without probe |
| Feature | Legacy HMT140 | Current RFL100 System |
|---|---|---|
| Wireless technology | Direct 2.4 GHz Wi-Fi | VaiNet wireless through AP10 |
| Primary measurements | Model-dependent temperature, RH, analog or contact inputs | Temperature, RH and CO₂ using compatible probes |
| Network connection | Each logger connects to the Wi-Fi network | AP10 connects the VaiNet devices to the network |
| Local measurement memory | 3,060 samples | 30 days of local measurement storage |
| Product status | Discontinued | Current wireless viewLinc platform |
Seacom Process Instruments supports existing Vaisala HMT140 installations in Malaysia and supplies current Vaisala RFL100, AP10 and VDL200 monitoring hardware for pharmaceutical facilities, laboratories, hospitals, blood banks, warehouses, cold rooms and other controlled environments.
Our team can assist with installed-model identification, replacement selection, monitoring-point planning, VaiNet coverage assessment, network architecture, viewLinc integration, installation, commissioning, calibration planning and GxP validation documentation.
Contact Seacom with the HMT140 model number, measured parameters, probe details, device quantities, facility layout, network constraints and validation requirements for a suitable migration plan.